Preparing Your Laboratory
36
Preparing the Laboratory
The following sections describe the laboratory requirements in detail for the Clarus 500 GC.
Environmental Conditions
The laboratory in which the Clarus 500 GC is located must meet the following conditions:
•
A corrosive-free environment.
•
The instrument will operate with a laboratory temperature between 10 and 35 °C (50- 95
°F). For optimum instrument preformance, the room temperature should be controlled at
20°
±
2 °C.
•
The environment should be relatively dust-free to avoid sample and instrument
contamination problems.
•
Free of excessive vibration.
The Clarus 500 GC has
been designed for indoor use. Do not use the instrument in an area
where explosion hazards may exist.
Exhaust Vent Requirements
Exhaust venting is important for the following reasons.
•
It protects laboratory personnel from toxic vapors that may be produced by some
samples.
•
It helps to protect the instrument from corrosive vapors that may originate from the
sample(s).
•
It removes dissipated heat produced by the instrument and power supply.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......