Installing the
Clarus 500 GC
48
Remove the Encoder Protector Block
The Clarus 500 GC autosampler has been shipped with an encoder protector block. This is a
piece of antistatic polyethylene foam that protects the tower disk during shipment. Before you
install the autosampler tower onto your Clarus 500 GC, you
must
remove the encoder
protector block from the tower disk assembly.
To remove the encoder protector block:
1.
Carefully remove the autosampler tower from the shipping box.
2.
The piece of foam on top of the tower is the encoder protector block. Carefully slide
the block straight out from the tower disk.
CAUTION
Do not twist the block off. This will damage the tower disk.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......