Clarus 500 GC Installation Guide
55
Install the Syringe Cover
To install the syringe, proceed as follows:
1.
Before you install the syringe, move the autosampler tower so it faces the
front of the instrument; this is called the “Park" position (see Figure 6).
Figure 6. The autosampler tower in the “Park” position.
CAUTION
Always handle the syringe with extreme care. Be careful not to bend the
needle. Before installing the syringe, pull some solvent up into the barrel to
moisten the mechanism.
NOTE: Use only syringes sold by PerkinElmer for the Clarus 500 GC. Syringe plungers are not
interchangeable from one syringe to another.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......