Clarus 500 GC Installation Guide
25
Leak Testing
You are obligated under U.S. federal and state regulations to make certain that the ECD cell
is wipe-tested for leakage of radioactive materials at intervals of no longer than six months,
and that the analysis of these wipe tests is conducted by a person specifically licensed to do
so, either by the U.S. Nuclear Regulatory Commission or by an Agreement State. The
analyses can be performed by the firm listed below:
National Leak Test Center
P.O. Box 486
North Tonawanda, New York 14120
ECD Cell Failure or Damage
If a leak test detects more than 0.005
μ
Ci (microcurie) of removable radioactive material on
the surface of an ECD cell, or if the cell itself is damaged in such a way as to indicate that it
may no longer adequately shield the radioactive material inside, you must immediately
suspend operation of your chromatograph until the cell has been repaired or disposed of by
a
person specifically licensed to do so
. Any such incident must be reported by you to the
Regional Office, Inspection and Enforcement, U.S. Nuclear Regulatory Commission.
Reporting Radiation Incidents, Theft or Loss
Please read Regulation 10 C.F.R. Section 20.2201 and 20.2202. These describe your duties
should the radioactive material (Ni 63) in the ECD cell be lost, stolen, or released, or should
any person be exposed to radiation.
Other ECD Requirements
Regulation 10 C.F.R. Section 31.5 (see Appendix I) does not permit you to abandon the ECD
cell or export it. It may not be transferred except to a person specifically licensed to receive
it. Within thirty days of such a transfer, you must report to the Director of Nuclear Material
Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, D.C. 20555, the
name and address of the transferee. However, no report is needed to transfer a defective ECD
cell to PerkinElmer in order to obtain a replacement.
You may transfer the ECD cell to another general licensee, like yourself, only when it
remains at the same location to which it was shipped by PerkinElmer. Give the transferee a
copy of these instructions and the regulations in Appendix I, and report to the commission as
required in Regulation C.F.R. Section 31.5.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......