PPC Restrictor Information
110
Restrictors for Detector Gas Control
NOTE:
Maintain an inlet pressure between 60 and 90 psig.
Detector Gas
Restrictor
Number
Nominal Flow
Required (mL/min)
FID Air
6 450
FID Hydrogen
4 45
NPD Air
5 100
NPD Hydrogen 1 2
FPD Air
5 90
FPD Hydrogen
4 75
ECD Argon/Methane
4 30
ECD Nitrogen 4 30
TCD Helium 4 30
TCD Hydrogen 3 30
ElCD Hydrogen 3 25