In Countries with 230V Single-Phase Power
WARNING
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the
Clarus 500 GC.
1.
Select the appropriate plug and connect it to the three power cord wires as follows:
•
Connect the brown wire to line (AC).
•
Connect the blue wire to neutral (ACC).
•
Connect the green/yellow wire to chassis ground.
2.
Remove the three screws that secure the right side panel to the Clarus 500 GC. Then
remove the panel.
3.
Verify that the brown wire from the line filter is connected to E1, that the blue wire
from the line filter is connected to E2, and that the yellow/green wire is connected to
E3 on the AC Distribution P.C. board (see Figure 23).
4.
Make sure that a 10-A Slow-Blow fuse is installed in the F3 fuse holder and a wire is
soldered across the contacts of the F3 fuse holder to short this connection.
5.
Replace the right side panel and connect the power cord to a suitable electrical outlet.
In Countries with 230V Two-Phase Power
WARNING
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the
Clarus 500 GC.
1.
Select the appropriate plug and connect it to the three power cord wires as follows:
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......