Safety Practices
16
CAUTION
THERMAL RUNAWAY PROTECTION: The Clarus 500 GC
software shuts down the instrument if any heated zone exceeds 470
°C. Should this occur, the following error message is displayed
:
INSTRUMENT SHUTDOWN
xxx THERM RUNAWAY where xxx is the heated zone
Call your PerkinElmer Representative.
Instrument shutdown also occurs if there is a PRT (Platinum
Resistance Thermometer) or MPU (Micro Processor Unit) failure.
In these cases the following error message is displayed:
INSTRUMENT SHUTDOWN
xxx PRT ERROR where xxx is the failed zone.
Call your PerkinElmer Representative.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......