Connecting the Accessories
104
Table 3. READY and START Connections at TB1
Connection Function
READY OUT
TB1-10 (C), 11 (NO), 12 (NC)
Instrument READY OUT Relay:
These contacts are
used to tell an external device that the Clarus 500 GC
is ready. The normally open contact (NO) is closed
in the Ready state.
START OUT
TB1-7 (C), 8 (NO), 9 (NC)
Instrument START OUT Relay:
These contacts are
used to start an external device, such as an integrator,
when the Clarus 500 GC starts a chromatographic
run. The normally open contact is closed for 1 sec
when a run is started.
EXT. READY
(TB1-5, 6)
TB1-6 is signal
TB1-5 is ground
External Ready In:
The Clarus 500 GC requires that
these contacts be shorted together to become ready
and is shipped with a link across them. When using
an external device, such as an integrator, remove the
link and wire the device to provide a contact closure
indicating the Ready state. This will prevent the
instrument from becoming ready before the external
device is ready.
EXT. START
(TB1-3, 4)
TB1-4 is signal
TB1-3 is ground
External Start In:
Shorting these contacts will cause
the Clarus 500 GC to start a chromatographic run. It
is equivalent to pressing the RUN key.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......