DETAILED SETTINGS
Two-point Scaling
Measurement is performed at two positions and offset values are set for those mea-
surements. Both an overall offset can be set and the range can be changed.
This section describes how to set two-point scaling, using an example of correcting dis-
play values to match actual distances.
Example: Correcting Display Values to Match Actual Distances
Separate the two specified points by at least 1% of the rated measurement distance for the con-
nected Sensor.
For example, the rated measurement distance for the ZX-ED01T Sensor is 1 mm. Therefore, the
two specified points must be separated by 10
µ
m min.
Setting the First Point
1.
Set the first point by following steps
1.
to
15.
of the one-point scaling procedure.
p. 92
The range for two-point scaling is automatically
set based on the values entered for the two
points. Inverted display settings are ignored.
Sensor Head
Actual distance: 1.0 mm
Value to display: 0.8 mm
Sensing
object
Display value (mm)
Actual meas-
ured value (mm)
Slide gauge
Actual distance: 0.2 mm
Value to display: 0.3 mm
Sensing
object
Z166-E1-02A.book Page 95 Friday, June 3, 2005 4:18 PM
Summary of Contents for ZX-E -
Page 4: ...2 Preface ZX E Operation Manual Preface ...
Page 8: ...6 Preface Contents ZX E Operation Manual Preface ...
Page 11: ...9 ZX E Operation Manual Preface PREFACE ...
Page 12: ...10 Preface ZX E Operation Manual PREFACE ...
Page 17: ...Section 1 FEATURES 15 ZX E Operation Manual Section 1 FEATURES ZX E Features 16 ...
Page 24: ...22 Section 1 ZX E Features ZX E Operation Manual Section 1 FEATURES ...
Page 62: ...60 Section 3 Adjusting Linearity ZX E Operation Manual Section 3 BASIC OPERATION ...
Page 162: ...160 Appendices Quick Reference for Displays ZX E Operation Manual APPENDICES ...