DETAILED SETTINGS
Linear Output
Output Settings (Monitor Focus)
Linear output refers to the conversion of measurement results to a 3 to 21-mA current
output or a
−
5 to 5-V voltage output. This section describes how to choose either cur-
rent or voltage output and how to set the linear output range. Match the settings to suit
the connected external device.
Enter the output values for any two current values or voltage values to set the output
range.
Example:
Setting 0.2 mm to 4 mA Output and 0.8 mm to 20 mA Output (for Current Output)
Separate the two specified points by at least 1% of the rated measurement distance for the con-
nected Sensor.
For example, the rated measurement distance for the ZX-ED01T Sensor is 1 mm. Therefore, the
two specified points must be separated by 10
µ
m min.
Using the Zero Reset Function
Zero reset is released when monitor focus is set. Execute the zero reset again after setting monitor
focus.
Zero Reset, p. 112
Measured
value (mm)
Output current (mA)
Measured
value (mm)
Output current (mA)
Z166-E1-02A.book Page 103 Friday, June 3, 2005 4:18 PM
Summary of Contents for ZX-E -
Page 4: ...2 Preface ZX E Operation Manual Preface ...
Page 8: ...6 Preface Contents ZX E Operation Manual Preface ...
Page 11: ...9 ZX E Operation Manual Preface PREFACE ...
Page 12: ...10 Preface ZX E Operation Manual PREFACE ...
Page 17: ...Section 1 FEATURES 15 ZX E Operation Manual Section 1 FEATURES ZX E Features 16 ...
Page 24: ...22 Section 1 ZX E Features ZX E Operation Manual Section 1 FEATURES ...
Page 62: ...60 Section 3 Adjusting Linearity ZX E Operation Manual Section 3 BASIC OPERATION ...
Page 162: ...160 Appendices Quick Reference for Displays ZX E Operation Manual APPENDICES ...