ZX-E
Operation Manual
Section 5
DETAILED SETTINGS
Comparing Measured Values
(Previous Value Comparisons)
Use the previous value comparison function to ignore gradual changes in measured values
over time, due to factors such as temperature drift, and only detect and judge sudden
changes.
The hold function must be set before previous value comparison can be set. The difference
from the previous hold value with a PASS judgement becomes the measured value. For
example, if the judgement for that previous measurement is HIGH or LOW, the comparison is
performed with the hold value before that.
The hysteresis width setting will be disabled if the previous value comparison function is used.
Hysteresis Setting, p. 102
Moving to FUN mode and SPCL
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Hold value
Sudden change
Measured value
(after comparison
with previous val-
ue) X(n)-X(n-1)
Gradual change
HIGH threshold
Z166-E1-02A.book Page 88 Friday, June 3, 2005 4:18 PM
Summary of Contents for ZX-E -
Page 4: ...2 Preface ZX E Operation Manual Preface ...
Page 8: ...6 Preface Contents ZX E Operation Manual Preface ...
Page 11: ...9 ZX E Operation Manual Preface PREFACE ...
Page 12: ...10 Preface ZX E Operation Manual PREFACE ...
Page 17: ...Section 1 FEATURES 15 ZX E Operation Manual Section 1 FEATURES ZX E Features 16 ...
Page 24: ...22 Section 1 ZX E Features ZX E Operation Manual Section 1 FEATURES ...
Page 62: ...60 Section 3 Adjusting Linearity ZX E Operation Manual Section 3 BASIC OPERATION ...
Page 162: ...160 Appendices Quick Reference for Displays ZX E Operation Manual APPENDICES ...