Section 5
DETAILED SETTINGS
Output Settings for Non-measurement
The linear output method for when a reset is input can be set.
For Hold Measurements
Even if KEEP is set, the output before the first hold value is obtained will be the same as CLAMP.
Moving to FUN Mode and SPCL
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT Keys to display
SPCL on the main display.
Moving to RESET
3.
Press the UP or DOWN Key.
The sub-display will flash.
4.
Use the UP and DOWN Keys to display ETC
or ALL.
5.
Press the ENT Key.
Selection
Outputs
Judgement outputs
Linear output
KEEP (default)
The status immediately before measurement is stopped is held and output.
CLAMP
All OFF
Outputs the set CLAMP value.
The following options are available.
• For current output: 3 to 21 mA or maximum
(approx. 23 mA)
• For voltage output:
−
5 to 5 V or maximum
(approx. 5.5 V)
Z166-E1-02A.book Page 110 Friday, June 3, 2005 4:18 PM
Summary of Contents for ZX-E -
Page 4: ...2 Preface ZX E Operation Manual Preface ...
Page 8: ...6 Preface Contents ZX E Operation Manual Preface ...
Page 11: ...9 ZX E Operation Manual Preface PREFACE ...
Page 12: ...10 Preface ZX E Operation Manual PREFACE ...
Page 17: ...Section 1 FEATURES 15 ZX E Operation Manual Section 1 FEATURES ZX E Features 16 ...
Page 24: ...22 Section 1 ZX E Features ZX E Operation Manual Section 1 FEATURES ...
Page 62: ...60 Section 3 Adjusting Linearity ZX E Operation Manual Section 3 BASIC OPERATION ...
Page 162: ...160 Appendices Quick Reference for Displays ZX E Operation Manual APPENDICES ...