Section 4
MAIN APPLICATIONS AND SETTING METHODS
Measuring Reference Samples
The height of the reference sample is measured using position teaching and the mea-
surement result is registered as the HIGH threshold value.
The registered value becomes the reference for the threshold value set in step
.
Refer to
Section 5 Detailed Settings
for details on settings.
Position Teaching, p. 99
The reference sample height can also be set to 0.
Using the Zero Reset Function, p. 127
Setting Tolerance Judgement Values
Refer to the HIGH threshold registered in step
and set the upper and lower limits
(HIGH and LOW thresholds) for a PASS (OK) judgement.
The HIGH, PASS, and LOW judgement results will be output based on the threshold
values set here.
Refer to
Section 5 Detailed Settings
for details on operation.
Inputting Threshold Values Directly, p. 98
Measurement result
Judgement
Measurement result > HIGH threshold
HIGH
LOW threshold
≤
measurement result
≤
HIGH threshold PASS
LOW threshold > measurement result
LOW
Z166-E1-02A.book Page 70 Friday, June 3, 2005 4:18 PM
Summary of Contents for ZX-E -
Page 4: ...2 Preface ZX E Operation Manual Preface ...
Page 8: ...6 Preface Contents ZX E Operation Manual Preface ...
Page 11: ...9 ZX E Operation Manual Preface PREFACE ...
Page 12: ...10 Preface ZX E Operation Manual PREFACE ...
Page 17: ...Section 1 FEATURES 15 ZX E Operation Manual Section 1 FEATURES ZX E Features 16 ...
Page 24: ...22 Section 1 ZX E Features ZX E Operation Manual Section 1 FEATURES ...
Page 62: ...60 Section 3 Adjusting Linearity ZX E Operation Manual Section 3 BASIC OPERATION ...
Page 162: ...160 Appendices Quick Reference for Displays ZX E Operation Manual APPENDICES ...