![OHAUS T72XW AM Instruction Manual Download Page 78](http://html1.mh-extra.com/html/ohaus/t72xw-am/t72xw-am_instruction-manual_3537195078.webp)
EN 70
7000 SERIES INDICATORS
Automatic The count and total values for all IDs are cleared after printing a
Target Table report.
Manual
Pressing the
CLEAR
key
after the Target Table report has
been printed clears the count and total values.
Disabled
Each record’s count and total values must be re-set to zero
individually, in the Target Table view in setup, as described under
View Table, below.
View Table
Selecting View Table in the Target Table screen displays the first record found in the Target Table. Each record
occupies a single screen. A blank screen indicates the ID Table is empty. The following functions are accessible
from this view:
Key
Function
Opens ID NEW screen, where a new ID can be set up.
(
CLEAR
)
Prompts Delete record? Press
ENTER
to confirm and delete the
displayed record,
LEFT
arrow to display a message, Delete all
records? Press
ENTER
to confirm and delete all records,
LEFT
arrow to return to the view without deleting.
UP and DOWN arrow keys
Move the view to the next or previous ID.
(
ENTER
/PRINT)
Sends the ID Table to all ports with a Reports assignment.
LEFT
arrow key
Exits the ID Table view.
The information displayed for each record depends on the Tolerance Type selected under Operation. In addition to
the information shown for Target Values, above, if Totalization is Enabled these screens also include a count (n:)
and Total value.
ID NEW Screen
The following options are available when setting up a new ID and when editing an existing ID. The record is
updated when the
LEFT
navigation key is pressed to exit the screen.
Parameter
Function
ID
Enter up to a 16-character ID to be used to recall the ID record.
Once the ID has been entered and the record saved, the ID
cannot be changed. If the ID is incorrect, the entire record must
be cleared and a new record created.
If the entered value is not unique, an error message displays
–
ID already exists. Press
ENTER
to clear the error and enter a
unique value for the ID.
Note: To exit the screen without creating a new ID, leave the ID
field blank and press
ENTER
.
Target
Target value, in the scale’s primary weight units. Not shown if
the tolerance type is selected as Weight Value.
-Tolerance or Under Limit
Tolerance value below the target that is acceptable or low limit
of acceptable weight.
+Tolerance or Over Limit
Tolerance value above the target that is acceptable or high limit
of acceptable weight.
Description
Alphanumeric description associated with this ID.
n
Indicates the number of times a weight value has been
accumulated for this ID.
Total
Indicates the total weight accumulated for this ID.
UP and DOWN arrows
Navigates between the labels on the setup page.
ENTER
Confirm entry and move to next parameter.
LEFT
arrow
Enters the new or edited ID record and exits the new/edit mode.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Summary of Contents for T72XW AM
Page 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...