Table 1-5: Switch 1 Functions
Switch
Functions
Notes
SW1-1
Metrology Security Switch
(legal for trade)
When in the ON position, this switch prohibits changes to
metrological parameters in setup. This must be ON for
“approved” applications
This is true even if the Scale
Approval parameter is selected
as “None” in setup.
SW1-2
Master Reset
Set in the ON position and power cycle to perform a master
reset of all data in the indicator
Set in the OFF position during normal operation.
When a Master Reset is
performed, set SW1-4 to ON to
reset metrological significant
data, such as scale calibration,
GEO code, etc.
SW1-3
Flash Software
Set in the ON position for software download
Set in the OFF position during normal operation
SW1-4
Reset Calibration
Set in the ON position to reset calibration during a master reset
Set in the OFF position to retain current calibration values
during a master reset
Works with switch SW1-2
SW1-5
Not used
SW1-6
Not used
When both SW1-2 and SW1-4 are positioned ON and AC power is applied to the indicator, a Master Reset function
will be initiated. This procedure will erase all programming in the indicator and return all settings back to factory
default values. This process is described in the T72XW Technical Manual, Chapter 4,
Service and Maintenance
.
SD Card Installation
The SD memory card can be used for additional storage in the Checkweighing and Counting applications, and
must be installed if the Vehicle Application is used. Figure 1-10 shows the installation of an SD card into the socket
on the edge of the T72XW main board.
Note:
The indicator comes with a SD card pre-installed.
Figure 1-10: Sliding an SD Card into the SD Socket (left); SD Card Installed (right)
Capacity Label Instructions
The regulations in some locations require that the scale capacity and increment be shown on the front of the
indicator, near the display. To satisfy this requirement, a blue capacity label is included with the indicator that must
be completed and adhered to the front overlay.
The capacity label (shown in Figure 1-11) provides space for the Max, min, and e information for both ranges for
which the scale is programmed. If only one range is used, the unused portion of the label may be cut off with
scissors. Written information must be legible and a minimum of 2mm or 0.08 in. in height. A permanent marker
should be used for this information.
Figure 1-11: Preparing the Capacity Label
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Summary of Contents for T72XW AM
Page 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Page 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...