Checking and Adjustment
6
Pulse Laser Station NPL-302 Series Instruction Manual
173
X1
Face-1 X axis tilt value
Y1
Face-1 Y axis tilt value
When you have taken the measurement, the
message on the bottom line changes from
DO
NOT TOUCH
%
to
Turn to F2
.
2.
Take an F2 measurement to the same target.
Press
[ENT]
.
VA2
Face-2 vertical angle (tilt-off value)
HA2
Face-2 horizontal angle (tilt-off value)
X2
Face-2 X axis tilt value
Y2
Face-2 Y axis tilt value
When the observation on F2 is completed, four parameters are displayed.
3.
Do one of the following:
–
To return to the first observation screen, press
[ESC]
or the
Redo
softkey.
–
To set parameters on the instrument, press
[ENT]
or the
OK
softkey.
The NPL-332 has vertical axis adjustment only.
The header displays Tilt instead of X and Y.
Otherwise, the calibration procedure is the same.
Summary of Contents for NPL-302 Series
Page 4: ......
Page 5: ...Safety 1 In this chapter Q Introduction Q Laser Safety Q Warnings and Cautions ...
Page 12: ...Safety xii Pulse Laser Station NPL 302 Series Instruction Manual ...
Page 28: ...1 Introduction 8 Pulse Laser Station NPL 302 Series Instruction Manual ...
Page 46: ...2 Preparation 26 Pulse Laser Station NPL 302 Series Instruction Manual ...
Page 82: ...3 Getting Started 62 Pulse Laser Station NPL 302 Series Instruction Manual ...
Page 188: ...5 Menu Key 168 Pulse Laser Station NPL 302 Series Instruction Manual ...
Page 196: ...6 Checking and Adjustment 176 Pulse Laser Station NPL 302 Series Instruction Manual ...
Page 208: ...8 System Diagrams 188 Pulse Laser Station NPL 302 Series Instruction Manual ...