BASIC MEASUREMENT
29
General Voltage Setting
F2 (Speed) key
to select refresh
rate
DCV:
Press the F1 to F5 key to select the desired rate
Press the F6 (More 1/2) key for next page with
more options as the figure shown below.
ACV:
Press the F1 to F3 key to select the desired rate
F3 (Auto Zero)
key to enable
Auto Zero (DCV
mode only)
Background
Autozero provides the most
accurate measurements, but requires
additional time to perform the zero
measurement. With autozero
enabled (On), the DM2571
internally measures the offset
following each measurement. It then
subtracts that measurement from
the preceding reading. This prevents
offset voltages present on the
DM2571
’s input circuitry from
affecting measurement accuracy.
With autozero disabled (Off), the
DM2571 measures the offset once
and subtracts the offset from all
subsequent measurements.
Display
When turning On the Auto Zero, the display
shows an icon
indicating the Auto
Zero mode is currently being activated.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......