DM2571 User Manual
278
MEASure:FRESistance?
Returns the 4-wire resistance measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS:FRES?
>
+1.1806
Sets measurement mode to 4-wire resistance to auto range and return
measurement value.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
MEASure:FREQuency?
Returns the frequency measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS:FREQ?
> +0.2105
Returns the frequency measurement value as 21.5 kHz.
MEASure:PERiod?
Returns the period measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS:PER? MAX
Returns the period measurement value at the maximum range.
MEASure:CAPacitance
Returns the capcitance measurement value on the 1
st
display
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS:CAP?
Returns the capcitance measurement value.
MEASure:CONTinuity?
Returns the continuity measurement value on the 1
st
display.
Example: MEAS:CONT?
Returns the continuity measurement value.
MEASure:DIODe?
Returns the diode measurement value on the 1
st
display.
Example: MEAS:DIOD?
Returns the diode measurement value.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......