DM2571 User Manual
270
CONFigure:RESistance
Sets measurement to 2-wire Resistance on the 1
st
display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example:
CONF:RES 10e3,MIN
Sets the measurement mode to 2-wire Resistance with a 10k
Ω
range at the
lowest resolution.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
CONFigure:FRESistance
Sets measurement to 4-wire Resistance on the 1
st
display and specifies the
range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF:FRES 1e3,MAX
Sets the measurement mode to 4-wire Resistance with a range of 1k
Ω
at the
maximum resolution.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
CONFigure:FREQuency
Sets measurement to Frequency on the 1
st
display and specifies range.
Parameter:
[None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF:FREQ MIN:MAX
Sets the frequency range to max.
CONFigure:PERiod
Sets measurement to Period on the 1
st
display and specifies range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF:PER AUTO,MAX
Sets the measurement mode to period with a auto range.
CONFigure:CAPacitance
Sets measurement to Capcitance on the 1
st
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF),]
Example: CONF:CAP 10e-7
Sets the Capcitance range to 100nF.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......