ADVANCED MEASUREMENT
101
Compare
mode result
When the measured result is within the range of high and low
limit, the display shows as the figure below with purely black
background indicating the state of “Pass”.
However, when measured result is either above or less than
the limit range, the display appears as the figure below with
boldly red background indicating the state of “Fail”.
See the contents below for more details of each state in
compare mode
High If the compare result is High, the relative pins of
digital I/O port in action are as the follows.
Digital I/O: FAIL Out (Pin 6) and HIGH Limit
FAIL Out (Pin 7) are activated.
Low If the compare result is Low, the relative pins of
digital I/O port in action are as the follows.
Digital I/O: FAIL Out (Pin 6) and LOW Limit FAIL
Out (Pin 8) are activated.
Pass If the compare result is Pass, the relative pin of
digital I/O port in action is as the follows.
Digital I/O: PASS Out (Pin 5) is activated.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......