![NF DM2571 User Manual Download Page 337](http://html.mh-extra.com/html/nf/dm2571/dm2571_user-manual_1684902337.webp)
REMOTE CONTROL
333
Bit Name
Decimal
Definition
0
Operation
Complete
1
All commands prior to and including *OPC have been
executed.
1
Not Used
2
(Reserved for future use)
2
Query Error
4
The instrument tried to read the output buffer but it
was empty. Or, a new command line was received
before a previous query has been read. Or, both the
input and output buffers are full.
3
Device
Error
8
A device error, including a self-test error or calibration
error, occurred (an error in the -300 range or any
positive error has been generated).
4
Execution
Error
16
An execution error occurred (an error in the -200 range
has been generated).
5
Command
Error
32
A command syntax error occurred (an error in the -100
range has been generated).
6
Not Used
64
(Reserved for future use)
7
Power On
128
Power has been cycled since the last time the event
register was read or cleared.
The following table describes the Status Byte Register.
Bit Name
Decimal
Definition
0
Not Used
1
(Reserved for future use)
1
Not Used
2
(Reserved for future use)
2 Error Queue
4
One or more errors have been stored in the Error
Queue. Use SYST:ERR? to read and delete errors.
3
Questionabl
e Data
8
One or more bits are set in the Questionable Data
Register (bits must be enabled, see
STAT:QUES:ENAB).
4
Message
Available
16
Data is available in the instrument's output buffer.
5
Standard
Event
32
One or more bits are set in the Standard Event Register
(bits must be enabled, see *ESE).
6
Request
Service
64
One or more bits are set in the Status Byte Register and
may generate a Request for Service(RQS). Bits must be
enabled using *SRE.
7
Operation
Data
128
One or more bits are set in the Standard Operation
Register (bits must be enabled, see
STAT:OPER:ENAB).
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...