DM2571 User Manual
90
Background
Using the ACV or DCV measurement result, the
DM2571 calculates the dBm, dB or Watt value based on a
reference resistance value in the following way.
Equation
dBm
10 x log
10
(1000 x Vreading
2
/ Rref)
dB
dBm – dBmref
Watt
Vreading
2
/Rref
Parameters
Vreading
Input Voltage, ACV or DCV
Rref
Reference resistance simulating an output
load
dBmref
Reference dBm value
Measure dBm/Watt
Applicable to
Equation
dBm
10 x log
10
(1000 x Vreading
2
/ Rref)
Watt
Vreading
2
/Rref
Parameters
Vreading
Input Voltage, ACV or DCV
Rref (REF
Ω)
Reference resistance simulating an output load
Activate dBm
Press the Shift key + Math key to
activate Math setting menu as the
following figure shown.
Further press the F1 (Function) key
to enter the Math Function menu as
the figure shown below.
Press F3 (dBm) key to enable the
dBm function. The screen, after
activation, will appear as figure below.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......