REMOTE CONTROL
325
STATus:QUEStionable:CONDition?
Returns the contents of the Questionable Condition register.
Return parameter: <NR1>, Ex: +2
●A condition register continuously monitors the state of the instrument.
Condition register bits are updated in real time; they are neither latched nor
buffered.
●This register is read-only; bits are not cleared when read.
STATus:QUEStionable:ENABle
Set bits in the Ouesrionable Enable register.
Parameter: <NR1> (0 to 32767)
Example: STAT:QUES:ENAB 4099
Sets the bit0, bit1 and bit12 in Ouesrionable Enable register, 4099 = 2
0
+ 2
1
+
2
12
.
●The selected bits are then reported to the Status Byte. An enable register
defines which bits in the event register will be reported to the Status Byte
register group. You can write to or read from an enable register.
●A STATus:PRESet clears all bits in the enable register.
●The *PSC command controls whether the enable register is cleared at power
on.
STATus:QUEStionable:ENABle?
Returns the total number of the Ouesrionable Enable register.
Return parameter: <NR1>, Ex: +1
STATus:QUEStionable[:EVENt]?
Returns the total number of the Ouesrionable Event register.
Return parameter: <NR1>, Ex: +2
●An event register is a read-only register that latches events from the
condition register. While an event bit is set, subsequent events corresponding
to that bit are ignored.
●Once a bit is set, it remains set until cleared by reading the event register or
by sending *CLS (clear status).
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......