DM2571 User Manual
122
Compare
live-result in
MathDisp
The latest state of compare measurement
, whether it’s
“Pass”, “High” or “Low”, will also appear within each
mode of MathDisp. See the example below for the
“
High
”
result in Math+STAT mode.
The boldly red background along with the indicator
“HIGH” within the display means the compare result is
over the range of defined high limit.
Timing Diagram
for pins 5-8 when
the Compare
function is
activated
Low limit
High limit
P
P
P
F
F
H
L
Pin 5
PASS Out
Pin 6
FAIL Out
Pin 7 HIGH
Limit FAIL Out
Pin 8 LOW
Limit FAIL Out
Pin 5-8 Output
Compare On
Deactivate
Compare
measurement
To cancel the Compare measurement, press the
F1 (Function) key followed by clicking F1
(OFF) to deactivate or simply activate another
measurement.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......