![NF DM2571 User Manual Download Page 309](http://html.mh-extra.com/html/nf/dm2571/dm2571_user-manual_1684902309.webp)
REMOTE CONTROL
305
[SENSe:]TEMPerature:RESolution
Sets the Temperature measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:TEMP:RES MAX
Sets the Temperature resolution to the maximum.
[SENSe:]TEMPerature:RESolution?
Returns the temperature measurement resolution.
[SENSe:]TEMPerature:TRANsducer:TYPE
Sets the temperature probe type.
Parameter: [None] | TC | RTD | FRTD | THER | FTH
Example: SENS:TEMP:TRAN:TYPE RTD
Sets the temperature probe type to RTD.
[SENSe:]TEMPerature:TRANsducer:TYPE?
Returns the temperature probe type.
Return parameter: TC, RTD, FRTD, THER, FTH
[SENSe:]TEMPerature:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of Temperature
measurement.
Parameter: <NRf>(0 to 3600 s) | MIN | MAX | DEF
Example: SENS:TEMP:TRIG:DEL 0.001
Sets the trigger delay time to 1ms of Temperature measurement.
[SENSe:]TEMPerature:TRIGger:DELay?
Returns the trigger delay time in seconds of Temperature measurement.
Return parameter: <NRf>
[SENSe:]TEMPerature:ZERO:AUTO
Sets the auto zero mode to on, off or once of Temparature measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:TEMP:ZERO:AUTO OFF
Sets the auto zero to off.
[SENSe:]TEMPerature:ZERO:AUTO?
Returns the auto zero mode of Temparature measurement.
Return Parameter: 0 | 1, 1=ON, 0=OFF
[SENSe:]TEMPerature:RJUNction:SIMulated
Sets temperature simulation value of thermocouple measurement.
Parameter: <NRf> (-20.00 to 80.00) | MIN | MAX | DEF
Example: SENS:TEMP:RJUN:SIM 25.00
Sets the thermocouple junction temperature to 25
˚
C.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......