REMOTE CONTROL
331
The following table lists the bit definitions for the Questionable Data Register:
NOTE: The overload bits are set once per INITiate command. If you clear an
overload bit, it is not set again until a new INITiate is sent.
Bit Name
Decimal
Definition
0
Voltage
Overload
1
Only reported as event. In Conditon Register this bit
always returns 0. Read the Event Register.
1
Current
Overload
2
Only reported as event. In Conditon Register this bit
always returns 0. Read the Event Register.
2
Not Used
4
(Reserved for future use)
3
Not Used
8
(Reserved for future use)
4
Temperature
Overload
16
Only reported as event. In Conditon Register this bit
always returns 0. Read the Event Register.
5
Frequency
Overload
32
Only reported as event. In Conditon Register this bit
always returns 0. Read the Event Register.
6
Not Used
64
(Reserved for future use)
7
Not Used
128
(Reserved for future use)
8
Not Used
256
(Reserved for future use)
9
Resistance
Overload
512
Only reported as event. In Conditon Register this bit
always returns 0. Read the Event Register.
10 Capacitance
Overload
1024
Only reported as event. In Conditon Register this bit
always returns 0. Read the Event Register.
11 Lower Limit
Failed
2048
The most recent measurement failed the lower limit test.
12 Upper Limit
Failed
4096
The most recent measurement failed the upper limit
test.
13
Not Used
8192
(Reserved for future use)
14 Data Buffer
Overload
16384
Data buffer is full. One or more (oldest) measurements
have been lost.
15
Not Used
32768
(Reserved for future use)
The following table lists the bit definitions for the Operation Data Register:
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......