DUAL MEASUREMENT
67
Refresh Rate
Background
Refresh rate defines how frequently the DM2571
captures and updates measurement data. A faster refresh
rate yields a lower accuracy and resolution. A slower
refresh rate yields a higher accuracy and resolution.
Consider these tradeoffs when selecting the refresh rate.
Measurement Type Refresh Rate
DCV/DCI
5/s 20/s
60 (50)/s 100/s 400/s 1.2k/s 2.4k/s 4.8k/s 7.2k/s 10k/s
ACV/ACI
1/s 5/s
50/s
Frequency/Period
1s
100ms 10ms
Selection steps
1.
Toggle the active display between the 1ST
and 2ND display by pressing the Knob
key until click.
2.
Press the F2 (Speed) key to select a
desired rate for measurement. Press the
corresponding function key (F1
–
F5) in
accord with the desired option on screen
display. Also, press the F6 (More 1/2) key
to enter the next page with more options
when available.
3.
The refresh rate will be shown at the left
side of each display. See the figure below
shown.
1ST Display
Refresh Rate
2ND Display
Refresh Rate
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......