![NF DM2571 User Manual Download Page 305](http://html.mh-extra.com/html/nf/dm2571/dm2571_user-manual_1684902305.webp)
REMOTE CONTROL
301
[SENSe:]FREQuency:VOLTage:RANGe:AUTO
Sets the Frequency Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:FREQ:VOLT:RANG:AUTO ON
Turns the Auto-range on for the frequency measurement.
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?
Returns the Frequency Auto-range setting.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]PERiod:APERture
Sets the aperture time (gate time) for the period function(0.01s, 0.1s, 1s).
Parameter: <NRf> (0.01 | 0.1 | 1)
Example: SENS:PER:APER 0.1
Sets the gate time to 0.1 seconds for the period function.
[SENSe:]PERiod:APERture?
Returns the aperture time (gate time) for the period function.
Return parameter: <NRf>
[SENSe:]PERiod:CURRent:RANGe
Sets the frequency measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:PER:CURR:RANG MAX
Sets the period to the maximum range.
[SENSe:]PERiod:CURRent:RANGe?
Returns the period measurement range.
[SENSe:]PERiod:CURRent:RANGe:AUTO
Sets the Period Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:PER:CURR:RANG:AUTO OFF
Turns the Auto-range setting off for period measurement.
[SENSe:]PERiod:CURRent:RANGe:AUTO?
Returns the Period Auto-range setting.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]PERiod:INPutjack
Assigns an input port for the period function.
Parameter: <NR1> (0 | 1 | 2), 0=Voltage, 1=3A, 2=10A
Example: SENS:PER:INP 1
Sets the input jack to the current 3A input port.
[SENSe:]PERiod:INPutjack?
Returns the assigned input port used for the period function.
Return parameter: VOLT | 3A | 10A
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......