![NF DM2571 User Manual Download Page 278](http://html.mh-extra.com/html/nf/dm2571/dm2571_user-manual_1684902278.webp)
DM2571 User Manual
274
DATA:REMove? <reading_number>,[WAIT]
Reads and erases measurement values from reading memory up to the
specified <reading_number>.
The measurement values are read and erased from the reading memory starting
with the oldest measurement first.
Ex:SAMP:COUN 10
INIT
DATA :REM? 4
Returns:
-1.12816521E-04,-1.13148354E-04,-1.13485152E-04,-1.13365632E-04
●If you do not specify a value for <reading_number>, +9.9137 (Not
a Number) is returned.
●If reading_number is greater than the latest counts of measurement, it will
return the error. However, it will return data if reading_number of counts of
measurement reach the set threshold only when WAIT parameter is specified.
●The R? and DATA:REMove? queries can be used during a long series of
readings to periodically remove readings from memory that would normally
cause the reading memory to overflow. R? does not wait for all readings to
complete. It sends the readings that are complete at the time the instrument
receives the command.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......