GETTING STARTED
17
Fan Vents
For heat ventilation when machine
is under operation.
Sense LO Terminal
Accepts LO sense line in 4W
resistance measurement. For
details, see page 38.
Sense HI Terminal
Accepts HI sense line in 4W
resistance measurement. For
details, see page 38.
Input LO Terminal
Accepts ground (COM) line in all
measurements except the sense
line in 4W Resistance (page 38).
The maximum withstand voltage
between this terminal and earth is
500Vpk.
Input HI
Terminal
Used as an input port for all
measurements except for DC/AC
Current measurements.
DC/AC 3A Terminal
DC/AC current input
DC: 100μA
to 3A
AC:
100μA
to 3A
For details see page 34.
DC/AC 3.15A Input
Current Fuse
Holds the current fuse: T3.15A,
500V , 5*20mm
For fuse replacement details, see
page 336.