REMOTE CONTROL
259
Other Commands
ABORt
Aborts a measurement in progress, returning the instrument to the trigger idle
state.
●Use this to abort a measurement when the instrument is waiting for a trigger,
or for aborting a long measurement or series of measurements.
FETCh[X]?
Waits for measurements to complete and copies all available measurements to
the instrument's output buffer. The readings remain in reading memory.
X = null or 1 indicate 1
st
display value, X = 2 indicate 2
nd
display value
Example: SAMP:COUN 3
INIT
FETC?
Returns: -4.98748741E-01,-4.35163427E-01,-4.33118686E-01
●The FETCh? query does not erase measurements from the reading memory.
You can send the query multiple times to retrieve the same data.
●You can store up to 10,000 measurements in the reading memory of the
DM2571. If reading memory overflows,new measurements overwrite the
oldest measurements stored; the most recent measurements are always
preserved. No error is generated, but the Reading Mem Ovfl bit (bit 14) is set
in the Questionable Data Register's condition register.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......