DM2571 User Manual
280
Secondary Display: MEASure2 Commands
MEASure2[:VOLTage]:DC?
Returns the DC voltage measurement value on the 2
nd
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS2:VOLT:DC? 1,MIN
> +4.88519457E-04
Returns the DC voltage measurement value as 0.000488 V.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
MEASure2[:VOLTage]:AC?
Returns the AC voltage measurement value on the 2
nd
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS2:VOLT:AC? MIN
> +5.11895142E-04
Returns the AC voltage measurement value as 0.5118 mV.
MEASure2:CURRent[:DC]?
Returns the DC current measurement value on the 2
nd
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS2:CURR:DC? 1E-4
> -1.05580457E-07
Returns the DC current measurement value as -0.1055 μA.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
MEASure2:CURRent:AC?
Returns the AC current measurement value on the 2
nd
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: MEAS2:CURR:AC?
>
+2.20387154E-07
Returns the AC current measurement value as 0.2203 μA.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......