DM2571 User Manual
30
F4 (Input R) key
to select input
resistance
Background
Specify the input impedance to the
test leads (Input R). This specifies
the measurement terminal input
impedance, which is either Auto or
10 M
Ω.
The Auto mode selects high
impedance (Hi-Z) for the 100
mV, 1 V and 10 V ranges, and 10
M
Ω for the 100 V and 1000 V
ranges. In most situations, 10 MΩ is
high enough to not load most
circuits, but low enough to make
readings stable for high impedance
circuits. It also leads to readings with
less noise than the (Hi-Z) option,
which is included for situations
where the 10 M
Ω load is
significant.
Vs = ideal voltage of DUT
Rs = input impedance of DUT
Ri = input impedance of DM2571
(either 10M or 10G available (Hi-Z))
Deviation (%) = Rs/(Rs+Ri) * 100
Display
When “Auto” is selected, the display shows
an icon
indicating the Auto mode is
currently being activated.
Summary of Contents for DM2571
Page 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Page 2: ......
Page 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 76: ...DM2571 User Manual 72 ...
Page 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Page 176: ...DM2571 User Manual 172 ...
Page 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Page 244: ...DM2571 User Manual 240 ...
Page 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Page 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Page 362: ......
Page 363: ......
Page 364: ......
Page 366: ......
Page 368: ......