National Instruments Corporation
67
NI 4065 Calibration Procedure
14. Call the niDMM Cal Adjust Gain VI.
15. Output 100 k
from the calibrator with 2-wire compensation turned
off, but with external sense turned on.
Allow the calibrator output to
settle before proceeding
.
16. Call the niDMM Cal Adjust Gain VI.
17. Output 10 k
from the calibrator with 2-wire compensation and
external sense turned on.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
1 M
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 1 M
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100 k
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 100 k