National Instruments Corporation
49
NI 4065 Calibration Procedure
You have completed verifying DC current for the NI 4065. Select one of the
following options:
•
If you want to continue verifying other modes, go to the
section.
•
If you do
not
want to verify any additional modes
and
you are
performing a
pre-adjustment
verification, call the niDMM Close VI or
the
niDMM_close
function to close the session.
Verifying AC Current
To verify AC current for the NI 4065, complete the following steps:
1.
Reset the calibrator.
2.
Connect the NI 4065 and the Fluke 5700A/5720A calibrator using the
Fluke 5440 cable, as shown in Figure 7. Table 12 lists the cable
connections.
3.
Call the niDMM Reset VI to reset the NI 4065 to a known state.
4.
Set the current output on the calibrator to
NORM
and output 0 A. Allow
the calibrator to settle before proceeding.
9
1
1.0
10
3
–2.2
11
3
0
12
3
2.2
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_close
with the following
parameters
Instrument_Handle
: The instrument
handle from
niDMM_init
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the following
parameter:
Instrument_Handle
: The instrument
handle from
niDMM_init
Table 13.
DC Current Settings (Continued)
Iteration
Range (A)
Calibrator Output (A)