NI 4065 Calibration Procedure
64
ni.com
Adjusting 2-Wire Resistance
To adjust 2-wire resistance for the NI 4065, complete the following steps:
1.
Connect the NI 4065 and the Fluke 5700A/5720A calibrator using
two sets of Fluke 5440 cables, a double banana plug, and
two insulated, low EMF spade lugs, as shown in Figure 5. Table 9 lists
the cable connections.
2.
Wait two minutes for the thermal EMF to stabilize.
3.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
4.
Call the niDMM Control VI.
5.
Output 100 M
from the calibrator with 2-wire compensation and
external sense turned off.
Allow the calibrator output to settle before
proceeding
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
:
Cal Session
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_2_WIRE_RES
Range
:
100 M
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
:
Cal Session
Control Action
:
Commit