NI 4065 Calibration Procedure
32
ni.com
4.
Call the niDMM Reset VI.
5.
Refer to Table 8 for the appropriate calibrator output and function
parameter values as you complete the following steps:
a.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance
of this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the following
parameter:
Instrument_Handle
: The instrument
handle from
niDMM_init
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_4_WIRE_RES
Range
: The
Range
listed in Table 8 for
the current iteration