NI 4065 Calibration Procedure
72
ni.com
34. Call the niDMM Cal Adjust Gain VI.
35. Reset the calibrator.
36. Plug in the insulated banana plug shorting bar across the HI and LO
banana plug connectors on the NI 4065.
37. Wait two minutes for the thermal EMF to stabilize.
38. Call the niDMM Cal Adjust Offset VI.
39. Call the niDMM Cal Adjust Offset VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100 M
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10 M
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA