NI 4065 Calibration Procedure
86
ni.com
20 V
1 kHz
20
19.95 V
20.05 V
19.952 V
20.048 V
6 V
1 kHz
300
5.838 V
6.162 V
5.868 V
6.132 V
30 V
1 kHz
300
29.79 V
30.21 V
29.82 V
30.18 V
300 V
1 kHz
300
299.25 V
300.75 V
299.28 V
300.72 V
200 mV
10 Hz
200 m
195.9 mV
204.1 mV
196.92 mV
203.08 mV
200 mV
40 Hz
200 m
195.9 mV
204.1 mV
196.92 mV
203.08 mV
200 mV
20 kHz
200 m
199.5 mV
200.5 mV
199.52 mV
200.48 mV
200 mV
50 kHz
200 m
199.3 mV
200.7 mV
199.32 mV
200.68 mV
200 mV
100 kHz
200 m
196.84 mV
203.16 mV
196.84 mV
203.16 mV
2 V
10 Hz
2
1.959 V
2.041 V
1.9692 V
2.0308 V
2 V
40 Hz
2
1.959 V
2.041 V
1.9692 V
2.0308 V
2 V
20 kHz
2
1.995 V
2.005 V
1.9952 V
2.0048 V
2 V
50 kHz
2
1.993 V
2.007 V
1.9932 V
2.0068 V
2 V
100 kHz
2
1.9684 V
2.0316 V
1.9684 V
2.0316 V
20 V
10 Hz
20
19.59 V
20.41 V
19.692 V
20.308 V
20 V
40 Hz
20
19.59 V
20.41 V
19.692 V
20.308 V
20 V
20 kHz
20
19.95 V
20.05 V
19.952 V
20.048 V
20 V
50 kHz
20
19.93 V
20.07 V
19.932 V
20.068 V
20 V
100 kHz
20
19.684 V
20.316 V
19.684 V
20.316 V
200 V
10 Hz
300
195.85 V
204.15 V
196.88 V
203.12 V
200 V
40 Hz
300
195.85 V
204.15 V
196.88 V
203.12 V
200 V
20 kHz
300
199.45 V
200.55 V
199.48 V
200.52 V
200 V
50 kHz
300
199.25 V
200.75 V
199.28 V
200.72 V
200 V
100 kHz
300
196.76 V
203.24 V
196.76 V
203.24 V
Table 21.
AC Voltage Verification Limits (Continued)
Calibrator Output
Range (V)
1-Year Limits
24-Hour Limits
Amplitude
Frequency
Lower
Upper
Lower
Upper