NI 4065 Calibration Procedure
19. Call the niDMM Control VI.
20. Output 100 mV on the calibrator with the range locked to 2.2 V.
This range prevents a 50
calibrator output resistance from creating a
voltage divider with the internal resistance of the NI 4065.
Allow the
calibrator output to settle before proceeding
.
21. Call the niDMM Read VI.
a.
Subtract the previously stored 100 mV >10 G
mode offset from
this measurement, and verify that the result falls between the
limits listed in Table 20.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Control Action
:
Commit
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that the result falls
between the limits listed in Table 20
Maximum_Time
: –
1