NI 4065 Calibration Procedure
12
ni.com
a.
Store the measurement as the 100 mV >10 G
mode offset.
11. Set a writable niDMM property node to set the input resistance of the
NI 4065 to 10 M
.
12. Call the niDMM Read VI.
a.
Store the measurement as the 100 mV 10 M
mode offset.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeViReal64
with the following
parameters:
Instrument_Handle
: The
instrument handle from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPUT
_RESISTANCE
Attribute_Value
:
NIDMM_VAL_10_MEGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that this measurement
falls between the limits listed in Table 20
Maximum_Time
: –
1