NI 4065 Calibration Procedure
c.
Call the niDMM Cal Adjust Offset VI (for iterations 7, 8, 9, 10,
and 11 only).
6.
Repeat step 5 for each of the remaining iterations listed in Table 17.
7.
Reset the calibrator.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
: Set as shown in Table 17 for the
current iteration
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Table 17.
4-Wire Resistance Settings
Iteration
Calibrator
Output (
)
niDMM Cal Adjust Gain Parameters
Range
(
)
Expected Value
1
1 M
1 M
the display on the calibrator for 1 M
2
190 k
1 M
the display on the calibrator for 190 k
3
100 k
100 k
the display on the calibrator for 100 k
4
10 k
10 k
the display on the calibrator for 10 k
5
1 k
1 k
the display on the calibrator for 1 k
6
100
100
the display on the calibrator for 100
7
0
1 M
the display on the calibrator for 0
8
0
100 k
the display on the calibrator for 0
9
0
10 k
the display on the calibrator for 0
10
0
1 k
the display on the calibrator for 0
11
0
100
the display on the calibrator for 0