NI 4065 Calibration Procedure
62
ni.com
4.
Call the niDMM Control VI.
5.
Refer to Table 17 for the appropriate calibrator output and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed in the
Calibrator Output
column in Table 17 for the current iteration.
Allow the calibrator
output to settle before proceeding
.
Note
For all 4-wire measurements, external sense on the calibrator is turned on.
b.
Call the niDMM Cal Adjust Gain VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
:
Cal Session
Control Action
:
Commit
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
: Set as shown in Table 17 for the
current iteration
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: Set as shown in
Table 17 for the current iteration