National Instruments Corporation
19
NI 4065 Calibration Procedure
25. Set a writable niDMM property node to set the input resistance of the
NI 4065 to >10 G
.
26. Output –100 mV on the calibrator with the range locked to 2.2 V.
This range prevents a 50
calibrator output resistance from creating a
voltage divider with the internal resistance of the NI 4065.
Allow the
calibrator output to settle before proceeding
.
27. Call the niDMM Read VI.
a.
Subtract the previously stored 100 mV >10 G
mode offset from
this measurement, and verify that the result falls between the
limits listed in Table 20.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeVi
Real64
with the
following parameters:
Instrument_Handle
:
The instrument handle
from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPUT_
RESISTANCE
Attribute_Value
:
NIDMM_VAL_GREATER
_THAN_10_GIGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that the result falls
between the limits listed in Table 20
Maximum_Time
: –
1