NI 4065 Calibration Procedure
66
ni.com
10. Call the niDMM Cal Adjust Gain VI.
11. Output 1.9 M
from the calibrator with 2-wire compensation turned
off, but with external sense turned on.
Allow the calibrator output to
settle before proceeding
.
12. Call the niDMM Cal Adjust Gain VI.
13. Output 1 M
from the calibrator with 2-wire compensation turned off,
but with external sense turned on.
Allow the calibrator output to settle
before proceeding
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10 M
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 10 M
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10 M
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 1.9 M