NI 4065 Calibration Procedure
58
ni.com
2.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
3.
Call the niDMM Control VI.
4.
Refer to Table 16 for the appropriate calibrator output and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed in the
Calibrator Output
column in Table 16 for the current iteration.
Allow the calibrator
output to settle before proceeding
.
b.
Call the niDMM Cal Adjust Gain VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
:
Cal Session
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
300.0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_Control
with the following
parameters:
Instrument_Handle
:
Cal Session
Control Action
:
Commit
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_VOLTS
Range
: Set as shown in Table 16 for the
current iteration
Input_Resistance
: Set as shown in
Table 16 for the current iteration
Expected_Value
: Set as shown in
Table 16 for the current iteration