National Instruments Corporation
71
NI 4065 Calibration Procedure
30. Call the niDMM Cal Adjust Gain VI.
31. Output 0
from the calibrator with 2-wire compensation and external
sense turned on.
Allow the calibrator output to settle before
proceeding
.
32. Call the niDMM Cal Adjust Gain VI.
33. Call the niDMM Cal Adjust Gain VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100 k
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10 k
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
1 k
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The display on the
calibrator for 0