National Instruments Corporation
21
NI 4065 Calibration Procedure
30. Refer to Table 4 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance
of this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
b.
Set a writable niDMM property node.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
: The
Range
listed in Table 4 for
the current iteration
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeViReal64
with
the following parameters:
Instrument_Handle
: The
instrument handle from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPUT
_RESISTANCE
Attribute_Value
: The
Input
Resistance
value listed in Table 4
for the current iteration