NI 4065 Calibration Procedure
26
ni.com
e.
Call the niDMM Read VI.
f.
Verify that this measurement falls between the limits listed in
Table 21.
4.
Repeat step 3 for each of the remaining iterations shown in Table 5.
5.
Reset the calibrator.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that this measurement
falls between the limits listed in Table 21
Maximum_Time
: –
1
Table 5.
AC Voltage Linearity Settings
Iteration
Range (V
ac
)
Calibrator Output
Amplitude (V
ac
)
Frequency (kHz)
1
0.2
0.004
1
2
0.2
0.02
1
3
0.2
0.2
1
4
2
0.04
1
5
2
0.2
1
6
2
2
1
7
20
0.4
1
8
20
2
1
9
20
20
1
10
300
6
1
11
300
30
1
12
300
300
1