NI 4065 Calibration Procedure
16
ni.com
17. Set a writable niDMM property node to set the input resistance of the
NI 4065 to >10 G
.
18. Set a writable niDMM property node to set the aperture time and
number of averages of the NI 4065.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeVi
Real64
with the
following parameters:
Instrument_Handle
:
The instrument handle
from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPUT_
RESISTANCE
Attribute_Value
:
NIDMM_VAL_GREATER
_THAN_10_GIGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeViReal64
with the following
parameters:
Instrument_Handle
: The
instrument handle from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_APERTURE_
TIME
Attribute_Value
:
166.67 ms
(
200 ms
for
50 Hz PowerLine
Frequency)
Attribute_ID
:
NIDMM_ATTR_NUMBER_OF
_AVERAGES
Attribute_Value
:
3